4
Installing the 2100 Bioanalyzer System
Installing the 2100 Bioanalyzer System
Installing the 2100 Bioanalyzer System
After unpacking and checking the completeness of the shipment, the 2100
Bioanalyzer system is ready to be installed. Depending on your purchased item,
the installation requires up to two steps:
1 Setting up the PC.
2 Setting up the Agilent 2100 Bioanalyzer instrument.
If you have problems installing your instrument, refer to the Troubleshooting
NOT E
section of the Agilent 2100 Bioanalyzer System Maintenance and Troubleshooting
Guide.
Setting up your PC
For setting up the PC, please refer to the instructions supplied with the PC.
Setting up the 2100 Bioanalyzer Instrument
To set-up the 2100 Bioanalyzer instrument:
1 Locate bench space that meets the requirements described in
2 Place the instrument on the bench in an upright position.
Damage by electrostatic discharge
C AUTI ON
The heater plate and electrodes are sensitive to electrostatic discharge.
122
space"
on page 111.
Make sure that you are grounded before inserting a chip to prevent the
instrument from being damaged by electrostatic discharge. The lid is
grounded to leak off electrostatic energy of the operator.
"Bench