Site Requirements for the Agilent 2100 Bioanalyzer Instrument

Power Considerations

Power Considerations

The instrument's power supply has wide ranging capabilities and accepts any
line voltage in the range 100 - 240 V ±10% with a line frequency of 50 - 60 Hz ±5%
(see also
selector in the rear of the instrument.
Wrong voltage range, frequency or cabling
WARN IN G
Personal injury or damage to the instrument
110
Table 1
on page 115 for more details). Consequently, there is no voltage
Verify that the voltage range and frequency of your power distribution
matches to the power specification of the individual instrument.
Never use cables other than the ones supplied by Agilent Technologies
to ensure proper functionality and compliance with safety or EMC
regulations.
Make all connections to the unit before applying power.