VWR Avantor 612-6857 Manual De Instrucciones página 16

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

3. After unscrewing pull out the shaft.
(Fig. 30)
4. This is the piston-shaft sub-assembly.
(Fig. 31)
5. Autoclave the two sub-assemblies at 121°C and 15 psi
pressure for 10-15 mins.
(Fig. 32)
6. The volume adjustment knob should always be kept in the
'unlocked' position while autoclaving.
(Refer fig. 10.A)
Re-assembling after Autoclaving
1. Insert the piston completely into the cylinder and then reassemble the
instrument use in the calibration tool by screwing back the piston.
(Fig. 33)
2. Snap back the cap to complete the assembly.
(Fig. 34)
3. Dispenser is now ready for use.
(Fig. 35)
Re-calibation is required after autoclaving.
16. vwr.com I Instruction manual VWR
Bottle Top Dispenser with Recirculation Valve
®
Fig.30
Fig.31
Fig.32
Fig.33
Fig.34
Fig.35
Tabla de contenido
loading

Tabla de contenido