Dis-assembling and Assembling the Delivery Pipe
1. Unscrew the chuck nut and pull out the delivery pipe.
(Fig. 24) & (Fig. 25)
2. Clean the delivery pipe with deionized water.
3. First push the delivery pipe into the lower housing till
it stops going in further. (Fig. 26)
4. Screw the chuck nut to complete the assembly.
(Fig. 27)
Autoclaving
This instrument is autoclavable at 121° C (250° F) 1 bar absolute (15 psi) with a holding time of
at least 15 minutes.
Note-Only the piston needs to be removed for autoclaving the instrument.
Dis-assembling for Autoclaving
1. Pull the cap outwards to expose the Calibration Nut.
(Fig. 28)
2. Unscrew the Calibration Nut with the help of calibration tool
to dis-assemble the Piston and shaft out of the main housing.
(Fig. 29)