12
STANDBY
When the unit is in Low, Medium, High, or Ambient mode, press the Standby
button to place the unit in Standby mode.
When you select the Standby mode, the following events occur:
• Standby indicator light illuminates.
• Blower and heater are turned OFF.
• Alphanumeric display deactivates.
• Alarm and fault detection functions remain active.
• Operating timer pauses.
TEMP IN RANGE INDICATOR LIGHT
The Temp in Range indicator light illuminates when the temperature at the
blanket end of the hose is ±1.5˚C of the selected setting; this indicator light
does not illuminate in the Ambient mode.
FAULT INDICATOR LIGHT
When a system fault occurs, the amber Fault indicator light flashes and an
alarm sounds.
Please refer to What to Do in Case of a Fault Condition on page 15 for
additional information.
-
OVER
TEMP INDICATOR LIGHT
If the unit senses an over-temperature condition, the red Over-temp indicator
light flashes and an alarm sounds.
Please refer to What to Do in Case of an Over-Temperature Condition on
page 15 for additional information.
Bair Hugger Model 750 Temperature Management Unit Operator's Manual