Technical Data
EMD Technical Data
Battery-operated Blood Pressure Monitor
Medical Electrical Equipment needs special precautions regarding
EMD and needs to be installed and put into service according to the
EMD information provided in the following.
Portable and mobile RF communication equipment (e.g. cell phones)
can affect Medical Electrical Equipment.
The use of accessories and cables other than those specified may
result in increased emissions or decreased immunity of the unit.
Table 1 - EMISSION Limits -
Phenomenon
Conducted and radiated RF
EMISSION CISPR 11
Table 2 - IMMUNITY TEST LEVELS : Enclosure Port -
Phenomenon
Electrostatic discharge
IEC 61000-4-2
Radiated RF EM fields
IEC 61000-4-3
Proximity fields from RF wireless
communications equipment
IEC 61000-4-3
Rated power frequency magnetic
fields
IEC 61000-4-8
E-23
Compliance
Group 1, Class B
IMMUNITY TEST LEVELS
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
10 V/m
80 MHz - 2.7 GHz
80 % AM at 1 kHz
See table 3
30 A/m
50 Hz or 60 Hz