XIII - ELECTROMAGNETIC COMPATIBILITY
Important:
The power cord, the micro-motor cord, the ultrasonic handpiece cords and the multifunction
footswitch cord must be kept away from each other.
The I
C
requires special precautions to be taken with regard to electromagnetic
MPLANT
ENTER
compatibility.
It must be installed and prepared for use as described in chapter 4.
Certain types of mobile telecommunication devices such as mobile telephones are likely to interfere
with the I
C
MPLANT
ENTER
The recommended separation distances in this paragraph must therefore be complied with.
The I
C
must not be used near or on top of another device.
MPLANT
ENTER
If this cannot be avoided, its operation under the conditions of use must be checked beforehand.
The use of accessories other than those specified or sold by Satelec® as replacement parts may have
the consequence of increasing the emissions or decreasing the immunity of the I
13. 1 ELECTROMAGNETIC EMISSIONS
The I
C
is intended for use in the electromagnetic environment specified in the table below.
MPLANT
ENTER
The user and/or installer must ensure that the I
Emission test
RF emission - CISPR 11.
RF emission - CISPR 11.
Harmonic current emission
IEC 61000-3-2.
Voltage fluctuation and flicker
IEC 61000-3-3.
All manuals and user guides at all-guides.com
.
Compliance
Group 1
Class B
Class A
Complies
C
is used in such an environment.
MPLANT
ENTER
Electromagnetic environment - guidance
I
C
uses RF energy for internal operation.
MPLANT
ENTER
Therefore, its radiofrequency emissions are very low and are not likely
to cause any interference in nearby equipment.
I
C
is suitable for use in all establishments, including
MPLANT
ENTER
domestic establishments and those directly connected to the public
low voltage power supply network that supplies buildings used for
domestic purposes.
C
.
MPLANT
ENTER
23