Regular monitoring of the I
necessary in order to detect any problem.
It is important to keep the control unit
ventilation vents clean in order to avoid
abnormal heating.
11. 5 MICRO-MOTOR CORD AND SCALER CORD
MAINTENANCE
Important:
Do not use an abrasive product to clean the
cords.
The cords must not be immersed.
The cords must not be disinfected using
products containing acetone, chlorine or
bleach.
The cords must not be cleaned in an ultrasonic
bath.
a) Cleaning and disinfection
The cords must always be cleaned, disinfected
and sterilized after each procedure, using
alcohol, disinfectant agents or disinfectant wipes
for dental practices, such as S
b) Sterilization:
The cords can be sterilized in an autoclave under
the following conditions:
- Autoclave: Class B.
- Sterilization temperature: 134 °C.
- Time at sterilization temperature: 18 minutes.
11. 6 I-SURGE MICRO-MOTOR MAINTENANCE
Important:
Do not use an abrasive product to clean the I-
S
micro-motor.
URGE
The I-S
micro-motor must not be immersed.
URGE
The I-S
micro-motor must not be
URGE
disinfected using products containing acetone,
chlorine or bleach.
All manuals and user guides at all-guides.com
C
unit is
MPLANT
ENTER
™ wipes.
EPTOL
The I-S
micro-motor must not be cleaned in
URGE
an ultrasonic bath.
The I-S
micro-motor must not be
URGE
disassembled.
The I-S
micro-motor is a high-technology
URGE
device. To maintain optimal performance, the
following recommendations must be complied
with.
a) Cleaning and disinfection:
Use
cleaning
and
compatible with the I-S
S
).
®
PRAYNET
When using a spray, spray the outside and inside
of the I-S
micro-motor moderately to expel
URGE
residues.
Clean and disinfect the surfaces using a clean,
lint-free cloth impregnated with the product.
b) Lubrication:
Lubricate before each sterilization or at least
twice a day.
Use lubricants compatible with the I-S
motor (such as L
UBRIFLUID
Place a clean cloth over the I-S
ventilation vents to absorb the surplus lubricant.
Inject the lubricant through the nose of the
micro-motor using an oilcan of lubricant (or a
spray) for one-half to one second.
c) Sterilization:
The I-S
micro-motor can be sterilized in an
URGE
autoclave under the following conditions:
- Autoclave: Class B.
- Sterilization temperature: 134 °C.
- Time at sterilization temperature: 18 minutes.
disinfection
products
micro-motor (such as
URGE
micro-
URGE
).
®
micro-motor
URGE
19