Display:
T.D.S.: Total Dissolved Solids
-
FULL: Not applicable since the unit is supplied without storage tank
-
WORK: In operation
-
POWER: Power is connected
-
SOURCE: Water is connected
-
FLUSH: Back flush
-
Now, the water will start to flow from the waste water outlet (14). After some time, also purified water flows
out of the pure water outlet (2), however, at a lower flow rate. Low pressure or temperatures reduce the pure
water volume. The pure water, produced in the first 2 hours, has to be discarded as it may contain residues of
the disinfectant liquid which has been used in the module for storage. After 2 hours, the water can be used.
Fig. 4: platinum line plus
Available spare parts: Please refer to www.aqua-medic.de.
The display shows the remaining salt content of the osmosis water. However, a TDS indicator of "00" does not
mean that the water is completely free of dissolved ingredients. The display is only for controlling the resin and
membrane. If the value rises, the resin has to be replaced in the demineralization filter. To do so, turn the unit
off, open the 10" housing with the appropriate wrench, remove the cartridge and refill it. Then close the 10"
housing firmly and restart the system.
5. Maintenance
-
Flushing: The membrane will be flushed automatically at regular intervals. The LED Flush will be lit then.
-
Prefilter: If the tap water is not extremely polluted, a yearly replacement of both prefilter cartridges will be
sufficient. If the water is higher polluted, the cartridges have to be changed more often (at the latest when
pure water production decreases).
-
Membrane: The membrane normally has a lifetime of several years. Afterwards, it has to be replaced.
Usually, the TDS measurement shows 99 if the membrane is broken.
Flushing valve 400: The flushing valve has a limited lifetime. In case of a decrease in adjusted waste water
-
flow, the valve is encrusted with calcium and can be cleaned using acid (e. g. vinegar). If the waste water
flow increases, the valve has to be replaced.
9
1.
Water inlet
2.
Pure water outlet
3.
Flush valve
4.
Module's screw cap
5.
Module incl. membrane
6.
Prefilter housing for fine filter
7.
Prefilter housing for activated
carbon filter
8.
Demineralization filter
9.
High-pressure pump
10. Control unit
11. Power supply
12. Magnetic valve
13. Low-pressure button
14. Waste water outlet
15. Magnetic valve
16. High-pressure button
17. Pressure gauge
18. Mounting bracket
19. TDS sensor