7.5
Pain when expressing
Expressing must not be painful. If it hurts, the following possible reasons must be eliminated:
• The vacuum is set too high. Adjust the settings so that they are comfortable for you (see Chapter 7.2).
• The nipples are not centered. Ensure that your nipples are in the middle of the breast shell.
• The funnel of the breast shell does not fit. It is too small or too big. The breast shell size is right if:
-
your nipples move freely in the funnel when expressing and track the rhythmical movements of the breastpump
(about 3 mm space between the funnel and the nipple).
-
none or only a little bit of the areola is located in the funnel.
-
the breast milk flows and the breast feels soft all over after the expressing.
• The breastpump with the pumpset is switched on before positioning the breast shell. As a result, the nipple is difficult
to center.
Order the correct size breast shell from Ardo, your specialist retailer, or your qualified consultant. Ardo has the following
breast shell sizes in stock: ø 22 / 26 / 28 / 31, and 36 mm (see fold-out page).
8
Troubleshooting
The breastpump barely sucks or does not suck at all
• Ensure that the pumpset has been correctly assembled and is not damaged (see instructions for use for the
"Pumpset and accessories").
• Check that the pumpset is connected firmly to the breastpump.
• Check whether the USB cable and the power adapter are correctly attached and connected to the mains power.
If these steps do not resolve the problem, please contact the point of sale, the hire center, or the customer service team at
the manufacturer Ardo medical AG.