EN-48
The Dynamic Configurations are defined below (defaults in Bold).
Configure Item
Dynamic Mode
Operation Type
Start Weight
Auto Tare
Accumulate
Find Quality Products Online at:
Option(Bold is default)
Count down/Continuous
Manual/
Semi-Automatic/
Automatic
3.000 (with current unit)
Off/On
Off/Manual/Automatic
GlobalTestSupply
www.
Defender 5000 Indicators
Description
Count down:
There is a countdown time.
Continuous:
There is no countdown time.
Manual:
Place load on the pan.
Press Softkey
and countdown.
When the countdown is over, the average weight is
displayed.
Before doing next weighing, press Softkey "Reset".
Semi-Automatic:
Before weighing, the reading should be zero (Gross or
Net).
Place load (>=Start Weight) on the pan, the dynamic
weighing and countdown start automatically.
When the countdown is over, the average weight is
displayed.
Before doing next weighing, press Softkey "Reset".
Automatic:
Before weighing, the reading should be zero (Gross or
Net).
Place load (>=Start Weight) on the pan, the dynamic
weighing and countdown start automatically.
When the countdown is over, the average weight is
displayed.
After the load has been removed, the average weight
still displays until the duration time is over.
If the Softkey "Reset" is pressed, the average weight
will be cleared immediately.
Note:
During the weighing countdown, if a weight error (under
load / over load) is occurred, the weighing process will
stop immediately.
If "Auto Tare" is on, a container (weight >=5d) must
place on the pan first. After the terminal has done tare
automatically, place load on the pan to start dynamic
weighing.
If the operation type is Semi-Automatic/
Automatic, this menu will be shown. If the load bigger
than the weight, then the dynamic weighing will start.
Off: Auto tare is turn off.
On: The first stable weight (>=5d) will be tared as the
container weight.
Off: The Softkey "∑" doesn't display.
Manual: The Softkey "∑" displays. User can press the
key to accumulate the stable weight.
Automatic: The Softkey "∑" displays. The weight will
be accumulated automatically.
.com
to start the dynamic weighing