Linearization - KERN CFS Serie Manual De Instrucciones

Balanza de conteo de piezas / sistema de cálculo
Ocultar thumbs Ver también para CFS Serie:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 792

15 Linearization

Linearity shows the greatest deviation of a weight display on the scale to the value
of the respective test weight according to plus and minus over the entire weighing
range.
If linearity deviation is discovered during a monitoring of test resources, you can
improve this by means of linearization.
• Carrying out linearization is restricted to specialist staff possessing well
acquainted with the workings of weighing scales.
• The adjustment weights to be used must be adapted to the weighing scale's
specifications; see chapter 3.4 "testing instruments control".
• Provide for required adjustment weights; cf. subsequent table 1 or table 2.
• Observe stable environmental conditions. Stabilisation requires a certain
warm-up time.
• After successful linearisation you will have to carry out calibration;
see chapter 3.4 "testing instruments control".
Access to menu:
 Switch-on balance and during the selftest press
 Use the numeric keys to input password „9999" on display „Pin"
 Confirm input by
Table 1: Required adjustment weights KERN CFS
Max
300 g
3 kg
6 kg
15 kg
30 kg
50 kg
CFS/CCS-BA-e-1523
.
1.
50 g
100 g
0.5 kg
1 kg
2 kg
6 kg
5 kg
15 kg
10 kg
30 kg
15 kg
30 kg
.
2.
3.
200 g
2 kg
-
-
-
50 kg
4.
300 g
3 kg
-
-
-
-
65
Tabla de contenido
loading

Tabla de contenido