1. Overview
The Druck DPI620 Genii is a battery-powered instrument for electrical measure and source
operations and HART
user interface functions for all optional items. The touch-screen displays up to six different
parameters.
2. Equipment in the Box
The following items are supplied with the Druck DPI 620 Genii:
•
DC power supply/battery charger unit.
•
Li-polymer battery.
•
Set of six test leads.
•
AC Probe.
•
Quick Start Guide.
•
Stylus.
3. Optional Items
The items that follow are optional items which can be used with the Druck DPI 620 Genii:
•
Pressure Module Carrier, MC 620, this attaches directly to the Druck DPI 620 Genii to make
a fully integrated pressure instrument.
•
Pressure Module, PM 620, this attaches to the pressure module carrier (MC 620) or a
pressure station (PV 62X) to enhance the pressure measurement functionality.
•
Pressure Stations, PV 62X, if the Druck DPI 620 Genii is installed in a pressure station, it
becomes a fully integrated pressure calibrator.
3.1 Electrical Warnings
•
To prevent electrical shocks or damage to the instrument, do not connect more than 30V
CAT I between the terminals, or between the terminals and the ground (earth).
•
External circuits should have appropriate insulation to the mains.
•
To prevent electrical shocks, use only the Druck specified AC probe (Part: IO620-AC) to
measure AC voltages that are more than 20 Vrms. Do not connect more than 300V CAT II
between the IO620-AC leads, or between the leads and the ground (earth). Attach it to the
specified connections only.
•
This instrument uses a Lithium-Polymer (Li-Polymer) battery pack. To prevent an explosion
or fire, do not short circuit, do not disassemble, and keep it safe from damage.
•
To prevent an explosion or fire, use only the Druck specified battery (Part: IO620-
BATTERY), power supply (Part: IO620-PSU) and battery charger (Part: IO620-CHARGER).
•
To prevent battery leakage or heat generation, only use the battery charger and power
supply in the temperature range 0 to 40°C (32 to 104°F).
•
The power supply input range is 100 – 240Vac, 50 to 60Hz, 250mA, installation category
CAT II.
•
Position the power supply so not to obstruct the supply disconnecting device.
•
To make sure the display shows the correct data, disconnect the test leads before power is
set to on or change to another measure or source function.
•
To prevent a dangerous release of pressure, make sure that all the related pipes, hoses and
equipment have the correct pressure rating, are safe to use and are correctly attached.
communications. The Druck DPI 620 Genii also supplies the power and
®
Copyright 2013 Baker Hughes Company.
English–DPI 620G Safety Instructions | 1