5.3 Connecting ADCC

Preparation of the Afinion™ AS100 Analyzer
The Afinion™ AS100 Analyzer should have SW 5.00 or higher installed. The SW version installed on the Afinion™ AS100 Analyzer will appear in
the upper left corner of the screen during the Automatic start-up procedure. Contact your local Afinion™ Supplier for latest the Afinion™ AS100
Analyzer SW version.
Connect the ADCC unit to the Afinion™ AS100 Analyzer serial port with the ADCC RS232 cable. Connect the ADCC Ethernet port to the HIS/
LIS system network with an Ethernet cable. Power the ADCC by connecting the ADCCs USB cable to the Afinion™ AS100 Analyzer USB port and
switch ON the Afinion™ AS100 Analyzer. The ADCC will now automatically transfer the results from the Afinion™ AS100 Analyzer to the HIS/LIS
systems
Allow the Afinion™ AS100 Analyzer to perform the Automatic start-up procedure.
Adjusting time and date
Precaution for ADCC with SW 1.00 installed: The correct time and date should be set on the Afinion™ AS100 Analyzer
before connecting to the ADCC. Later time adjustments should be done prior to running any tests that day. When returning
to standard time from daylight saving time, allow a minimum of one hour from the last test run before adjusting the analyzer
time.
All manuals and user guides at all-guides.com
Configuration
Insert the IP address of the receiving system [0-255 0-255 0-255 1-254]
[0-65535] (0 = not set)
ASTM Low Level: For relevant information, see chapter 11. The communication protocol is based
on ASTM E 1381 - 95
ASTM High Level: For relevant information, see chapter 11. The communication protocol is based
on ASTM E 1394 - 97
HL7: For relevant information, see chapter 11. The communication protocol is based on HL7
version 2.4
New Results Only: Only test results obtained after first connection between ADCC and Afinion™
AS100 Analyzer will be transferred to the EPR system
All Results: All results stored in the Afinion™ AS100 Analyzer memory and future test results
obtained after first connection between ADCC and Afinion™ AS100 Analyzer will be transferred
to the EPR system
If the ADCC has been connected to the Analyzer before, only results after last connection
will be transferred. To send all results again the Clear device list must be used. See the category
Reset
Only Patient: Only Patient results will be transferred to the EPR system
Patient and QC: Patient and Control results will be transferred to the EPR system
Configuration
The length of name may consist of up to 30 characters. Invalid characters are \ | & ^ ~
For relevant information, see chapter 11
The length of name may consist of up to 30 characters. Invalid characters are \ | & ^ ~
For relevant information, see chapter 11
Visit Number: For relevant information, see chapter 11
The length of the password may consist of up to 15 characters
New password
Confirm new password
✓ Restore factory settings. Password and Device list will not be cleared
✓ It is possible to switch the ADCC to another Afinion™ AS100 Analyzer without changing the
settings in the box. Choosing the «Clear Device list» will delete this memory
«Update settings» will confirm and store new settings. The settings will be stored in a non volatile
memory