It is recommended to access to this menu only after a complete charge of the batteries. From the
main menu choose "Advanced" , then "Set-Up" and finally "Battery info" .
Six codes will be visualized with the following meanings:
COD1 = 0 expected voltage threshold reached.
COD1 = 1 max. charge time reached.
COD2 = value of the battery voltage at the start of the charge.
COD3 = value of the battery voltage at the end of the charge.
COD4 = charge duration (from 1 to 840 minutes, ideal time 720 minutes).
COD5 = charger/power supply unit connection duration.
COD6 = Battery pack voltage value.
According to the values above described, it is recommended to replace the battery when COD1
= 1 and COD3 < 7 ,4 volts. Or when COD3-COD2 >= 2 volts and COD4 <600. Or when COD6 is
lower than 5,8 volts.
Furthermore it is recommended to replace the battery pack after 3 months in which the device
has not been used. After that period, batteries generally lose their ability to charge making the
recharge dangerous.

ACCESSORIES

Use and storage of the electrodes and the cables.
- After using the multi-purpose single patient and/or single-use electrodes, they must be stored
using their plastic film and placed in the plastic bag.
- Avoid that the electrodes touch each other or that they overlie one over the other.
- Once the package has been opened, the electrodes can be used for 25-30 applications.
- The electrode must be always used with clean hands and they must be replaced if they are not
perfectly in contact with the skin.
- If using non self-adhesive electrodes it is suggested to clean the surface with proper cleansers
that respect the requirements described in the manual.
- The electrodes must be stored in their bag and in an environment that respects the requirements
described in the manual.
- After the end of a treatment, unplug the cables from the connectors and clean them carefully
with proper cleansers that respect the requirements described in the manual.
- After cleaning and drying them, they must be folded up and placed in the plastic bags supplied
with the cables.

DISPOSAL OF THE DEVICE

Do not throw the device or part of it in the fire, but dispose of the product in the specialized
centers and respecting the directives in force within your country. When the product has to be
disposed of, the user can give it back to the retailer when purchasing a new unit.
A correct separate waste collection or following what above mentioned contribute to avoid possi-
ble negative effects on environment and health and promote the reuse and/or recycle of materials
of which the device is composed. The illegal disposal of the product by the user involves the
application of the administrative fines according to the current regulations.

INSTRUCTIONS FOR USE

For a correct connection of the device to the user, proceed as follows:
- Connect the cables to the outlets on the main unit.
- Connect the electrodes to the cables connected to the device.
- Place the electrodes on the skin.

TURNING ON

Turn on the Electrostimulator by pressing the ON/OFF/OK button for 3 seconds until a sound
alarm is heard.