Do not make current measurements of 10 A for longer than
30 seconds. Exceeding this time may cause damage to the
meter and/or test leads.
When the current of 10 A, the interval between successive
tests should be approx. 15 minutes.
5.9.1 The range of 10 A
To perform the current measurement:
set the rotary switch at 10A,
press MODE button to display the following values on the display:
AC, if you measuring Alternating Current,
DC, if you measuring Direct Current,
connect black test lead to COM terminal and red test lead to 10A
terminal,
disconnect the power supply from the tested circuit and then con-
nect the meter in series (for DC circuits: connect the red probe to
the positive pole),
turn on the power supply of the circuit,
read the measurement result on the display.
5.9.2 Range of mA, µA
To perform the current measurement:
set the rotary switch at
(it is recommended to start the measurement from the posi-
tion of the rotary switch in the higher range)
press MODE button to display the following values on the display:
AC, if you measuring Alternating Current,
DC, if you measuring Direct Current,
connect black test lead to COM terminal, and red test lead to
VΩHz%
disconnect the power supply from the tested circuit and then con-
nect the meter in series (for DC circuits: connect the red probe to
the positive pole),
turn on the power supply of the circuit,
read the measurement result on the display.
NOTE
mA or
mACAP terminal,
CMM-11 – USER MANUAL
μA