Connection A - High Vacuum Flange - Agilent Technologies TV 1001 SEM Manual De Instrucciones

Tabla de contenido
17
Technical Information
TV 1001 SEM Controller Installation
Connection A – High Vacuum Flange
Figure 45
It is mandatory to connect the pump in such a way to withstand the torque
WARNING!
specified in the "INSTRUCTION FOR USE" section. Specifically pay attention
to clamp design, material of clamps and bolts and bolt fixing torque.
To facilitate assembly and dismantling, apply Felpro C-100 high temperature
lubricant to the screw threads protruding from the flange and between the
nuts and flange.
Note that the connections can be made only with the bolt in the lower side.
Attach the units and tighten each one in turn. Repeat the sequential tightening
until the flange faces meet.
Exercise care when tightening nuts and bolts to avoid creating dents in the
CAUTION!
envelope as this may cause the pump rotor to lock.
284/296
For ISO-K flange connections, fix the two flanges with the clamps as shown
in the figure.
TV 1001 SEM User Manual / 87-900-973-01 (H)
Tabla de contenido
loading

Este manual también es adecuado para:

7698931869893286989469698978

Tabla de contenido