Switching on and Use of TV 1001 SEM

To switch on the TV 1001 SEM it is necessary to supply the mains.
The integrated controller automatically recognizes the interlock and
start signals presence and start up the pump.
The first pump start up is in "Soft Start" mode. When the start up
cycle is finished, the "Soft Start" mode automatically is disabled, and
the following start ups are without the "Soft Start" mode. To re-
enable the "Soft Start" mode it must be activated by the suitable
software command (see the paragraph "RS 232/485
COMMUNICATION DESCRIPTION" in the appendix "Technical
Information").
The green LED located on the TV 1001 SEM base front panel
indicates with its flashing frequency the system operating conditions:
TV 1001 SEM User Manual / 87-900-973-01 (H)
with no flashing: the pump is normally rotating;
slowly flashing (period of about 400 ms): the system is in ramp,
or in braking, or in Stop, or in "Waiting for interlock" status;
fast flashing (period of about 200 ms): error condition.