follow the instructions in the instruction manual.
TECHNICAL SPECIFICATIONS
RF6 Plus Nebulizer
In compliance with 93/42/EEC Directive
Minimum fill volume: 2 ml
Maximum fill volume: 8 ml
APPLIED PARTS
The applied parts of type BF are:
patient accessories (1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 14)
FLAEM
Operating
compressor
pressure
range
F700 / F1000
0.8 bar
F2000
1.2 bar
Data shown is in accordance with Flaem Nuova internal procedure I29-P07.5. (2) In vitro characterization
(1)
carried out at the Inamed Research GmbH Co. KG on behalf of TÜV Rheinland LGA Products GmbH – Germany in
compliance with the new European Standard for aerosol therapy units. Standard EN 13544-1. Further details are
available upon request.
Operating conditions:
Temperature: min 10°C; max 40°C
Air humidity: min 10%; max 95%
Atmospheric pressure: min 69KPa; max 106KPa
Storage conditions:
Temperature: min -25°C; max 70°C
Air humidity: min 10%; max 95%
Atmospheric pressure: min 69KPa; max 106KPa
BREATHABLE
MMAD
FRACTION
(μm)
(2)
< 5 μm (FPF) %
3.2
73
2.8
76
10
DELIVERY ml/min
(approx)
(2)
open valve-
holder MAX
0.42
0.54
(1)
closed valve-
holder MIN
0.26
0.31