TROUBLESHOOTING
If any abnormality arises during use, please check the following points:
POSSIBLE
SYMPTOMS
CAUSES
Unit does not turn
Batteries have
on when the START/
run down.
STOP
button is pushed.
Battery polarities
have been positioned
incorrectly.
EE measurement
Cuff has been placed
error symbol shown
incorrectly.
on display or the
Did you talk or
blood pressure
value is displayed
move during
excessively low
measurement?
(or high).
Shaking of the wrist
with the cuff on.
E1 error symbol
Air circuit
shown on display
abnormality.
E2 error symbol
Inflation pressure
shown on display
exceeding 300 mmHg.
E3 error symbol
Error determining
shown on display
measurement data.
EP error symbol
System error.
shown on display
Note: If the unit still does not work, contact HoMedics Consumer
Relations. Under no circumstance should you disassemble or
attempt to repair the unit by yourself. Contact information for
HoMedics Consumer Relations Department can be found on the
warranty page.
CORRECTION
Replace them
with two new "AAA"
alkaline batteries.
Re-insert the
batteries in the
correct positions.
Wrap the cuff properly
so that it is positioned
correctly. Measure again.
Keep wrist steady during
measurement. Measure
again.
Measure again.
Switch the unit off, then
measure again.
Measure again.
Turn off monitor and
measure again. If 'EP' error
still appears on
the display, please call
a Consumer Relations
representative.
29
SPECIFICATIONS
Power Source:
Two 1.5V DC "AAA" Alkaline batteries
Measurement Method: Oscillometric
Rated Range of Cuff
0-300 mmHg
Pressure:
Pressure: 40~280 millimeters Mercury (mmHg)
Measurement Range:
Pulse: 40~199 beats/minute
Rated Range of
40-280 mmHg
Determination:
Pressure: ±3 mmHg
Accuracy:
Pulse: ±5% of reading
Pressure Sensor:
Semi-conductor
Inflation:
Pump driven
Deflation:
Automatic pressure release valve
Memory Capacity:
60 memories
Auto-Shut-Off:
1 minute after last button operation
Battery Life:
Approx. 250 measurements
Operation
Temperature: 41°F~104°F (5°C~40°C)
Environment:
Humidity: 15% ~ 93% RH
Storage/Transportation
Temperature: -13°F~158°F (-25°C~70°C)
Environment:
Humidity: Less than 93% RH