TV 1001 CONTROLLER INSTALLATION

The controller can be mounted in two position:
bottom mounting (as per the complete system)
side mounting.
To install the controller execute the following pro-
cedures.
Bottom mounting
See the following figure.
1. Turn the pump upside-down;
2. Place the three fixing blocks with the through
holes toward the inside, and fix them with the
three M6 screws to the pump bottom;
3. Place the controller on the blocks, with the
pump cable toward the pump body;
4. Fix the controller to the blocks by means of the
three socket head screws M5;
5. Turn the pump again;
6. Plug the line card and connect the pump cable
to the pump;
7. Connect the mating connector with the jumper
on the interlock signal to start the pump.
Side mounting
The L-shaped bracket (P/N 969-9298) is available
as an option.
See the following figure.
1. Place the L-shaped bracket on the controller,
with the pump cable toward the bracket, and
fix it by means of the three socket head
screws M5, the washers and the nuts. The 6
holes on the bracket allows to mount the con-
troller with the rotation angle that you like;
2. Turn the pump upside-down;
3. Place the bracket on the pump bottom (the
hole on the bracket has the same diameter as
the pump bottom flange);
4. Screw a little bit the three screws M6 in the
holes on the pump bottom;
5. Rotate the bracket until the controller is on the
chosen position;
6. Tighten the screws properly;
7. Turn the pump again;
8. Plug the line card and connect the pump cable
to the pump;
9. Connect the mating connector with the jumper
on the interlock signal to start the pump.