Tv 1001 Pump With Standard Rack Controller; Water Cooling Kit Installation - Varian TV 1001 Navigator Manual De Instrucciones

Tabla de contenido
TV 1001 Pump with Standard Rack
Controller
When the TV 1001 pumps are used with the stan-
dard rack controller, it is necessary to utilize the air
cooling kit model 969-9315.
Fan specification:
− air flow: 55 l/s (120 CFM)
− input voltage: 120 Vac 50-60 Hz
− maximum power: 17 W
− dimensions: 127 x 127 mm (5 x 5 inches)
To fix the fan to the pump, position it ensuring that
the holes in the plate line up with those in the
pump base.
Insert the respective washers and screws and
tighten with a screwdriver.
Connect the fan to the controller.

WATER COOLING KIT INSTALLATION

Two types of water cooling kits are available to be
mounted when the pump is used under heavy load
conditions or when air cooling is insufficient.
The two model part numbers are: 969-9337
(metallic model), and 969-9347 (plastic model).
Model 969-9337
Model 969-9347
The items of the plastic model kit must be assem-
bled as shown in the following figure
84
TECHNICAL INFORMATION
CAUTION
87-900-945-01(G)
Tabla de contenido
loading

Este manual también es adecuado para:

Tv 551Tv 701969-8822

Tabla de contenido