Air Cooling Kit Installation; Tv 1001 With Navigator Controller - Varian TV 1001 Navigator Manual De Instrucciones

Tabla de contenido
Switch on the heater while the turbopump is in op-
eration. In the event of turbopump overheat, the
pump will be automatically cut out by the thermis-
tor sensor.
NOTE
The turbopump must be "baked" only when operat-
ing with an inlet pressure less than 10
with water cooling.
If the chamber of the system is "baked" at a high
temperature, a shield should be installed to pre-
vent thermal radiation heating the high vacuum
flange on the pump. The maximum temperature
allowed for the inlet flange is 120° C.

AIR COOLING KIT INSTALLATION

TV 1001 with Navigator Controller

An air cooling kit (mod. 969-9297) is available to
improve the TV 1001 cooling during heavy opera-
tional conditions (optional).
Fan specifications:
3
− air flow: 200m
/h
− input voltage: 24 Vdc
− dimensions: 127 x 127 x 38 mm (5 x 5 x 1.5 in.)
− power: 4.7 W
The fan bracket is shaped so that it can be
mounted close to the pump.
-4
mbar and
CAUTION
To fix the fan to the TV 1001 case execute the fol-
lowing procedure (see the following figure):
1. Fix the fan to the suitable bracket by means of
the furnished screws;
2. Fix the bracket to the pump body between the
pump and the controller;
3. Connect the fan supply to the P4 connector of
the controller.
83
TECHNICAL INFORMATION
87-900-945-01(G)
Tabla de contenido
loading

Este manual también es adecuado para:

Tv 551Tv 701969-8822

Tabla de contenido