authorised FLAEM service centre or to your trusted dealer.
• Use the device only in dust-free conditions, otherwise treatment could be compromised.
• Do not wash the device under running water or by immersion and keep it safe from being
sprayed by water or other liquids.
• Do not expose the device to particularly extreme temperatures.
• Do not place the device near sources of heat, in direct sunlight or in excessively hot
rooms.
• Do not obstruct or put objects into the filter or its related housing in the device.
• Never obstruct the air vents located on both sides of the device.
• Always use it on a rigid surface that is clear of obstacles.
• Make sure there is no material obstructing the air vents before each use.
• Do not put any objects in the air vents.
• Repairs, including the replacement of the supply cord, are to be carried out by FLAEM
authorised personnel only, by complying with the information provided by the
manufacturer.
• The average expected duration for the compressor series are: F400: 400 hours, F700: 700
hours, F1000: 1000 hours, F2000: 2000 hours.
• WARNING: Do not modify this device without authorisation from the manufacturer.
• The Manufacturer, the Vendor and the Importer shall be held responsible for safety,
reliability and performance only if: a) the device is used in compliance with the
instructions for use b) the wiring where the device is being used is in compliance with
safety regulations and current laws.
• Interactions: the materials used in contact with medication have been tested with a
vast range of medications. However, in view of the variety and continuous evolution of
pharmaceuticals, interactions cannot be ruled out. We recommend using the medication
as soon as possible once it has been opened and preventing prolonged exposure in the
nebuliser cup.
• The manufacturer must be contacted about any problems and/or unexpected events
concerning operation and for any clarifications on use, maintenance/cleaning.
• Interactions: The materials used in the medical device are biocompatible in accordance
with the provisions of Directive 93/42 EC and subsequent amendments. However, the
possibility of occurrence of allergic reactions cannot be entirely excluded.
• The amount of time required between preservation conditions and use is approximately
2 hours.
OPERATING INSTRUCTIONS
Before each use, clean hands thoroughly and clean the device as described in the
section on "CLEANING SANITISATION DISINFECTION". During use, it is advisable to
protect yourself from any dripping. It is recommended that each person use their
own nebulizer cup and accessories to prevent risk of infection due to contamination.
This device is suitable for the administration of medical substances and not, for
which the administration via aerosol is foreseen; these substances are to be in any
case prescribed by the Doctor. In case of too thick substances, the dilution with a suit-
able physiological solution could be needed, according to the medical prescription.
1. Plug the power cord (A6) into a power socket corresponding to the voltage of the unit.
This must be positioned so that it is not difficult to make the disconnection from the
mains.
2. Insert the nozzle (C1.2) in the upper part (C1.3) pressing as shown by the 2 arrows in the
"Connection diagram" in point C1. Insert the Speed selector (C1.4) in the upper part (C1.3)
as shown in the "Connection diagram" in point C1. Pour the medication prescribed by
the doctor into the lower part (C1.1). Close the nebuliser by turning the upper part (C1.3)
clockwise.
10
3.
4.
5.
6.
WA
H
Pr
in
of
ac
SO