i-SENS KetoSens BT Manual Del Usuario página 8

Sistema de monitoreo de b-cetona en sangre
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Limitations
Do not use for diagnosis of or screening for diabetes mellitus.
The system should not be used to test neonates.
Do not test samples other than fresh capillary whole blood
obtained from the fingertip.
Do not use the system at altitudes above 10,000 feet
(3,000 meters).
Do not use when hematocrit is outside the acceptable
hematocrit range for testing of 20 % to 55 %.
Severe dehydration (excessive water loss) may cause
inaccurate results. If you believe you are suffering from severe
dehydration, consult your healthcare professional immediately.
For in vitro diagnostic use only.
Critically ill patients should not be tested with this device.
Inaccurate results may occur in severely hypotensive
individuals or patients in shock.
Incorrect result may occur in individuals who are dehydrated.
The meter and lancing device are for single patient use. Do not
share these items with anyone, including other family members!
Do not use on multiple patients!
Do not reuse; each test strip is for single use only.
Do not use when humidity is higher than 90 % and lower than
10 %, as extremes in humidity may affect results.
For single patient use only.
For over-the-counter use.