Limitations
Do not use for diagnosis of or screening for diabetes mellitus.
•
The system should not be used to test neonates.
•
Do not test samples other than fresh capillary whole blood
•
obtained from the fingertip.
Do not use the system at altitudes above 10,000 feet
•
(3,000 meters).
Do not use when hematocrit is outside the acceptable
•
hematocrit range for testing of 20 % to 55 %.
Severe dehydration (excessive water loss) may cause
•
inaccurate results. If you believe you are suffering from severe
dehydration, consult your healthcare professional immediately.
For in vitro diagnostic use only.
•
Critically ill patients should not be tested with this device.
•
Inaccurate results may occur in severely hypotensive
•
individuals or patients in shock.
Incorrect result may occur in individuals who are dehydrated.
•
The meter and lancing device are for single patient use. Do not
•
share these items with anyone, including other family members!
Do not use on multiple patients!
Do not reuse; each test strip is for single use only.
•
Do not use when humidity is higher than 90 % and lower than
•
10 %, as extremes in humidity may affect results.
For single patient use only.
•
For over-the-counter use.
•