Understanding Error Messages

52 www.i-sens.com
A used test strip was inserted.
> Repeat the test with a new test strip.
The blood or control solution sample was
applied before the
> Repeat the test with a new test strip and
wait until the
applying the blood or control solution
sample.
The temperature during the test was above or
below the operating range.
> Move to an area where the temperature is
within the operating range 50–104 ˚F
(10–40 ˚C) and repeat the test after the
meter and test strips have reached a
temperature within the operating range.
The blood sample is of insufficient volume or
blood is too thick.
> Repeat the test using a new test strip.
This error message may appear when the
wrong blood β-Ketone test strip is used instead
of the KetoSens Blood β-Ketone test strip.
> Repeat the test after inserting a KetoSens
test strip.
symbol appeared.
symbol appears before