Step 3
Shake the KetoSens Control
Solution bottle well before each test.
Remove the cap and squeeze the
bottle to discard the first drop. Then
wipe the tip with a clean tissue or
cloth. Dispense a drop of control
solution onto a clean non-absorbent
surface. It helps to squeeze a drop onto the top of the cap as
shown. After the
symbol appears on the display, apply the
solution to the narrow edge of the test strip until the meter beeps.
Make sure the confirmation window fills completely.
Note
The meter may switch off if the control solution is not applied
within 2 minutes of the
symbol appearing on the screen. If the
meter turns off, remove the strip, reinsert, and start from step 2.
Step 4
A test result will appear after the meter counts
down from 8 to 1.
The result is stored in the meter's memory but it
is not included in the averages.
30 www.i-sens.com