BTL7-A/E501-M_ _ _ _-J-DEXC-TA12
Magnetostrictive Linear Position Sensor – Rod Style
9
Adjust with the calibration device (continued)
Set end point
► Bring magnet to the new end point.
► Hold down
for at least 2 s.
⇒ The new end point is set with the last
valid end value.
Adjust end value
► The end value can be changed using
gradient of the curve changes
(see page 20).
► End calibration procedure: Briefly press
simultaneously (< 1 s).
⇒ Set position value is saved.
End adjustment
► Briefly press
and
simultaneously (< 1 s).
⇒ The current position value is displayed once the
buttons are released.
Any of the individual steps for settings can be
selected. The adjustment process can be
ended at any time.
1
= blue button
= gray button
2
www.balluff.com
> 2 s
1
1
and
. The
1)
2
2
and
< 1 s
< 1 s
Type of protection "db" and "t"
Flameproof enclosure
Displayed values (example)
At 0...10 V
9.89 V
2
10.00 V
10.00 V
8.00 V
8.00 V
1
2
1
2
english
At 4...20 mA
19.13 mA
20.00 mA
20.00 mA
16.80 mA
16.80 mA
23