Toto Touch Control Unit Operating Guide
1. Start, stop and pause the control unit.
2. Air cell side selection - Left only, Right only or Both.
3. Air cell inflation time - Length of time the air cell remains inflated; minimum 30 minutes, maximum
240 minutes.
4. Air cell rest time - Length of time the air cells remain deflated and the platform is flat; minimum
30 minutes, maximum 240 minutes.
5. Alarm mute and interface lock mode activation.
Turning the Toto Touch Control Unit On
•
Press and hold the start/stop button (1) for three seconds.
•
A single beep confirms activation along with default lit LEDs.
•
Note the system default settings below; these can be changed as appropriate.
o
Air cell side selection - both sides.
o
Air cell inflation time - 120 minutes.
o
Air cell rest time - 120 minutes.
•
The pump automatically inflates the left side of the platform; this can be changed as appropriate.
INFLATE TIME
2
1
L & R
R
touch
3
L
INFLATE TIME
1
L & R
R
touch
L
4
240 min
180 min
30 min
120 min
60 min
60 min
120 min
30 min
180 min
240 min
REST TIME
240 min
180 min
30 min
120 min
60 min
60 min
120 min
30 min
180 min
240 min
REST TIME
13