Pressalit SCT 3000 Instrucciones De Uso Y Mantenimiento página 26

Camilla de ducha
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 156
Immunity to proximity fields from RF wireless communication
equipment
Test
standard
EN 61000-4-3*
en
Electrical fast transient/burst
Test
standard
EN 61000-4-4
26
The product must not be affected by radiating RF signals,
for example from a mobile phone or radio, that could
cause the product to take action or to make components
defective.
Immunity test level
385 MHz, 27 V/m, 18 Hz PM (50 % duty cycle square wave)
450 MHz, 28 V/m, FM +/- 5 kHz deviation, 1 kHz sine
710 MHz, 9 V/m, 217 Hz PM (50 % duty cycle square wave)
745 MHz, 9 V/m, 217 Hz PM (50 % duty cycle square wave)
780 MHz, 9 V/m, 217 Hz PM (50 % duty cycle square wave)
810 MHz, 28 V/m, 18 Hz PM (50 % duty cycle square wave)
870 MHz, 28 V/m, 18 Hz PM (50 % duty cycle square wave)
930 MHz, 28 V/m, 18 Hz PM (50 % duty cycle square wave)
1720 MHz, 28 V/m, 217 Hz PM (50 % duty cycle square wave)
1845 MHz, 28 V/m, 217 Hz PM (50 % duty cycle square wave)
1970 MHz, 28 V/m, 217 Hz PM (50 % duty cycle square wave)
2450 MHz, 28 V/m, 217 Hz PM (50 % duty cycle square wave)
5240 MHz, 9 V/m, 217 Hz PM (50 % duty cycle square wave)
5500 MHz, 9 V/m, 217 Hz PM (50 % duty cycle square wave)
5785 MHz, 9 V/m, 217 Hz PM (50 % duty cycle square wave)
The product must not be affected by rapid pulse signals
through the cord. This could cause the components to
become defective, or the product could initiate actions.
Immunity test level
+/- 1 kV , +/- 2 kV
100 kHz repetition frequency
Tabla de contenido
loading

Este manual también es adecuado para:

R8572313R8582313R8592313R85723188582318R8592318

Tabla de contenido