Operating The Pump - Agilent Technologies Turbo-V 250 Manual De Instrucciones

Tabla de contenido
17
Technical Information
Use

Operating the Pump

After the starting period, if the system has a vacuum leak or the
pressure in the pump/chamber is high (from 1 mbar to atmosphere),
the pump continues to operate indefinitely. If the gas load at the
turbopump inlet flange continues to stay high, the power drawn by
the turbopump increases up to the maximum value. Than the Turbo-V
pump is slowed down in proportion to the gas load at least until it
reaches 3 KRPM.
Even if any previous display selection was made. This will occur
either in NORMAL operation or with the LOW SPEED selected. As
soon as the gas load decreases, the pump will automatically
accelerate to reach normal operation. The pump can be stopped at
any rotational speed and can be restarted at any rotational speed
from either the front panel buttons or the remote connections. The
controller automatically synchronizes the output to the rotational
speed of the pump and then accelerates linearly up to the nominal
speed or within steps if the Soft Start has been selected.
228/240
Turbo-V 250 Controller User Manual / 87-900-856-01 (E)
Tabla de contenido
loading

Este manual también es adecuado para:

969-9504

Tabla de contenido