17
Technical Information
Optional Serial Port
Serial Communication Port
Communication serial port connections and mini- mum connection
configuration are shown in the following figures. The communication
port mating connector is supplied with the RS 232 PCB
(AMP/Cannon or equivalent 9-pin "D" type male connector). The
external cable (not supplied) between the host computer and the
controller doesn't require any crossed wires so that signals are
connected correctly. For example, the Transmit data signal from
controller (pin 2) must be connected to the host computer's Receive
data line (pin 2) and vice versa. Consult the host computer's
instruction manual for its serial port connections.
Agilent cannot guarantee compliance with FCC regulations for radiated
NOTE
emissions unless all external wiring is shielded, with the shield being
terminated to the metal shroud on the O-subconnector. The cable should be
secured to the connector with screws.
RS 232-422 Communication Descriptions
Communication format:
The baud rate is programmable via front panel from 600 to 9600
baud. The controller is factory-set for 9600 baud operation.
When a baud rate of 2400, 4800, 9600 baud is used, the requested information
NOTE
from to controller must be sent with a minimum interval of 1 second to avoid the
controller hang-up.
218/262
8 data bit
no parity
1 stop bit
Turbo-V 1000HT PCB Controller User Manual / 87-900-906-01
(F)