Enter ACTtrac
From the Main Menu:
1. Select [Cartridge Lot].
2. To navigate through the list, use the arrows on either side of the box.
When the ACTtrac
appear: [Add Lot Number] and [Exit to Main Menu].
1. To enter the serial number, select [Add Lot Number].
2. If the serial number has an "AT" at the beginning of it, press the "." key and the
"AT" will appear in the serial number field.
3. Enter the rest of the serial number using the numeric keypad.
4. To confirm the entry, press [Enter].
Manually Enter a Cartridge Lot Number and Expiration
Date
From the Main Menu:
1. Select [Cartridge Lot].
2. To navigate through the list, use the arrows on either side of the box.
When there are no lot numbers for a cartridge type, the following variable function
keys appear: [Add Lot Number], [Add Exp Date], and [Exit to Main Menu].
1. To enter the lot number, select [Add Lot Number].
2. Enter the cartridge lot number using the numeric keypad.
3. To confirm the entry, press [Enter].
4. To enter the expiration date for the selected lot number, select [Add Exp
Date]. The date format is yy-mm-dd.
5. Enter the cartridge expiration date.
6. To confirm the entry, press [Enter].
When a cartridge type has a single lot number entered and a second lot is to be
entered, the following variable function keys will appear: [Add Lot], [Remove Lot],
[Edit Lot/Exp Date] and [Exit to Main Menu].
1. To enter the second lot number, select [Add Lot].
2. Enter the cartridge lot number using the numeric keypad.
3. To confirm the entry, press [Enter].
4. To enter the expiration date for the selected lot number, select [Add Exp
Date]. The date format is yy-mm-dd.
5. Enter the cartridge expiration date.
6. To confirm the entry, press [Enter].