VASOVIEW HEMOPRO Power Supply

WARNINGS AND PRECAUTIONS

1.
Only use with the supplied regionally compatible Power Cord connected to a Hospital Grade receptacle.
2.
Use only with the VASOVIEW HEMOPRO Endoscopic Vessel Harvesting System. Use with any other
instrument may damage the instrument and/or the VASOVIEW HEMOPRO Power Supply and could
prevent proper functioning during use.
3.
Do not drop the Power Supply.
4.
Do not sterilize the Power Supply. Follow recommended cleaning instructions as described in these
Instructions for Use.
5.
Do not use in the presence of flammable materials (e.g., alcohol, flammable anesthetics).
6.
There are no risks associated with proper disposal of the Power Supply. Follow any local regulations
regarding proper disposal of used electronic equipment.
7.
Do not allow patient to contact grounded metal parts.
8.
The safety and effectiveness of the Power Supply has not been fully evaluated in patients with cardiac
pacemakers or metallic implants.
9.
Inspect the Power Supply and all accessories, connections, and cords before each use, looking for damage
and ensuring that casing is intact.
10.
Do not immerse the Power Supply in liquids.
11.
Check to make sure that there is no evidence of damage to the control knob on the Power Supply.
12.
Inspect the on/off switch and power cord for evidence of rough handling, abrasion, kinks, or other signs
of damage.
13.
Inspect all labels and markings for legibility.
14.
Check the condition of the housing, cover, handle, markings, and warning labels for evidence of damage.
15.
Call Customer Service (888-880-2874 or 408-635-6800) if you discover any damage or situation that
may jeopardize the safety and/or effectiveness of this device.
16.
The Power Supply has no serviceable parts.
HANDLING, STORAGE and CLEANING
1.
Do not drop. Store in a cool, dry place. Avoid prolonged exposure to extreme temperatures.
2.
Exterior of the VASOVIEW HEMOPRO Power Supply may be cleaned using a soft cloth moistened with
a solution of water and hospital grade detergent or disinfectant. Gently wipe down all cords, switches,
buttons, and device housing, taking special care not to get liquid into housing or electrical components.
Gently dry the entire Power Supply, including cords.
Table 1. Recommended Environmental Conditions
Transport
-15°C to 50°C
25 to 95
700hPa to 1060hPa
Storage
Operating
15°C to 30°C
15°C to 30°C
30 to 85
45 to 75
700hPa to 1060hPa
700hPa to 1060hPa
ENGLISH
3