VASOVIEW HEMOPRO Power Supply
ENGLISH
6
5
4
3
2
1
Figure 1: VASOVIEW HEMOPRO Power Supply
1. ON/OFF Switch
2. Extension Cable Connector
3. Power Cord Connector
4. Power Setting Knob
5. LED Power-On Indicator
6. Hanger
Read all information carefully before using.
Caution: Federal (USA) law restricts this device to sale by, or on the order of, a physician.
Important: These Instructions for Use are designed to assist in the use of the VASOVIEW HEMOPRO Power
Supply.
DEVICE DESCRIPTION
The VASOVIEW HEMOPRO Power Supply is a reusable, AC-powered unit intended to be used only with the
VASOVIEW HEMOPRO Endoscopic Vessel Harvesting System. The VASOVIEW HEMOPRO Power Supply
is intended to be connected to AC power via a grounded hospital grade power cord and features an on/
off switch and green power-on LED indicator. Output power is delivered to the VASOVIEW HEMOPRO
Endoscopic Vessel Harvesting System when the Activation Toggle on the Harvesting Tool is pulled into its
most proximal position. An intermittent tone indicates the Power Supply is activated. The level of power
delivered to the VASOVIEW HEMOPRO Harvesting Tool is set by a knob on the Power Supply front panel;
higher numbers indicate a higher level of energy. The Power Supply may be placed on a flat, non-sterile surface
adjacent to the sterile field, or may be suspended from a nearby IV pole.
Compliance with Standards. The VASOVIEW HEMOPRO Power Supply complies with IEC60601-1
(UL 2601-1 and CAN/CSA-C22.2 No. 601.1-M90) requirements for type Class I, Type CF equipment, Type
CF applied parts, and meets electromagnetic compatibility requirements of IEC-60601-1-2.
Electrical Requirements. Input: 100-240 VAC, 50/60 Hz, 80 W. Maximum Output: 32 W, 6 V D.C.
HOW SUPPLIED
The VASOVIEW HEMOPRO Power Supply is non-sterile. Contents of the VASOVIEW HEMOPRO Power
Supply: One (1) VASOVIEW HEMOPRO Power Supply device, One (1) power cord.
2