DISINFECTING
Two wipes must be used: one wipe to clean and a second wipe to disinfect.
After cleaning your meter or lancing device, repeat steps 1, 2, and 3 using a
second germicidal wipe to disinfect.
Allow the meter and lancing device to remain visibly wet for full contact time
detailed below.
Clorox
Germicidal Wipes
®
NOTE:
Your meter and lancing device should be CLEANED and DISINFECTED a
minimum of once per week and immediately when blood is visible, for up
to 2 cleaning and disinfection cycles per week over the 4-year use life of
the device.
Over time, cleaning and disinfecting may cause meter and lancing device
deterioration. This system has been validated to a 4-year use life by exposing
the meter and lancing device to 832 wipes (416 cycles) with the disinfectant
product listed on page 37 of this user manual. This supports cleaning and
disinfection twice per week over the 4-year use life of the device.
Contact Time
1 minute