Running And Stopping A Method - Varian 212-LC Manual Del Usuario

Tabla de contenido

Running and Stopping a Method

Press Activate. Enter the number of the method file you wish to run (00
– 25), press Enter and then press Run. This process will start the
method file at T0 conditions (times out after 60 minutes) until Run is
pressed again or the "Start"/"Common" pins on the back panel are
shorted (by remote contact closure Figure 2). Both of these commands
will initiate the method run starting at T0. Pressing Stop or shorting the
"Stop" Common" pins at any time will terminate the method run. If the
slave pump (Pump B) is not connected via the RS232 cable when Run is
pressed, a warning will be displayed which reads:
Make sure the RS232 cable is installed properly between the master and
slave pumps. You can override this warning by pressing Enter. In this
condition the master pump will operate by itself with a flow profile that
matches the solvent gradient profile for the Pump A.
17
*****Warning*****
Slave not connected
Hit enter to confirm
Tabla de contenido
loading

Tabla de contenido