"SILENT POWER"
The Silent Power high speed air driven turbines are class IIa medical devices intended to be fitted on dental
equipments, built in compliance with 93/42/EEC Directive and with IEC 601-1, EN 60601-1regulations.
Install such instruments only on units in conformity with the above said standard!
It is indispensable to read the instructions of this manual carefully before using these handpieces.
The high speed air driven turbine is intended to be used by Dentists only, who may avail themselves of the
assistance of authorized personnel.
It must be operated in environmental conditions in accordance with the prescriptions of the dental unit user's
manual.
The Manufacturer will in no case be held responsible for the safety, reliability and performance of
the device where:
-the premises do not satisfy the basic requirements (as described in the dental unit user's manual);
-the assembly, the fitting of additional devices, the calibration and any successive adjustments or repairs are not
carried out by qualified technicians authorized by Castellini S.p.A. and in possession of a valid professional
license;
-any unauthorized modifications, tampering or incorrect servicing are carried out or where non-original spare
parts and/or components are used;
-the device is not used in accordance with the operating instructions or is used for purposes other than its
intended use.
Speed
Air pressure (handpiece)
Air delivery
Water pressure
Burr stem diameter
Max burr lenght
Min burr lock lenght
Max burr diameter
Warning!
- High speed handpiece!
- An excessive pressure may damage the tooth!
- Cool the operating area conveniently! Stop the turbine immediately if the cooling media
don't flow!
- Do not use burrs with dimensions exceding the above mentioned limits! Do not use worn
or damaged burrs! Handle burrs with care wearing protective gloves!
- Only CE-marked burrs (or other similar tools) according to Directive 93/42/EEC must be
used.
- Before every use, check that the burr is correctly locked in the gripper (see section "BURR
LOCKING").
- The burr release button becomes very hot if it is pushed while the turbine is running. Pay
the greatest attention not to push it unintentionally against the Patient's oral cavity.
- Do not use the instrument if it is visibly damaged or if unusual noises and/or vibrations
are produced. Contact an authorised technical service centre.
- For the use of this instrument the rubber dam must be applied on the patient.
The high speed air turbine has to be subjected to intermittent operation: 20 min work, 10 min stand-by.
WARNING

TECHNICAL SPECIFICATIONS

Silent Power Gold
Silent Power Silver 4L
350.000 ÷ 366.000 rpm'
260 ÷ 280 kPa
(2.6 ÷ 2.8 bar)
48 ÷ 51 lt/min
70÷140 kPa (0.7÷1.4 bar)
1.590÷1.600 mm
(ISO 1797-1)
26 mm
11.7
2 mm
Pag. / Seite 9
Silent Power Gold
Miniature
380.000 ÷400.000 rpm
280 ÷ 300 kPa
(2.8 ÷ 3 bar)
44 ÷ 47 lt/min
70÷140 kPa (0.7÷1.4 bar)
1.590÷1.600 mm
(ISO 1797-1)
19 mm
11.2
2 mm
UK
Silent Power Silver 2
Silent Power Silver 2L
350.000 ÷ 366.000 rpm'
260 ÷ 270 kPa
(2.6 ÷ 2.7 bar)
48 ÷ 51 lt/min
70÷140 kPa (0.7÷1.4 bar)
1.590÷1.600 mm
(ISO 1797-1)
26 mm
11.7
2 mm