Hach NA5600 sc Na+ Manual De Instrucciones página 14

Ocultar thumbs Ver también para NA5600 sc Na+:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 62
2. Select an option.
Note: Select SET PARAMETER first, then SET FUNCTION and then ACTIVATION.
Option
ACTIVATION
SELECT
SOURCE
SET
PARAMETER
SET FUNCTION
SET TRANSFER
SET FILTER
SCALE 0mA/4mA Sets the analog output value range to 0–20 mA or 4–20 mA (default).
• LINEAR CONTROL function
Option
SET LOW VALUE Sets the low measurement value that is represented as 0 or 4 mA at the
SET HIGH VALUE Sets the high measurement value that is represented as 20 mA at the
• PID CONTROL function
Option
SET MODE
PHASE
SET SETPOINT Sets a measurement value as the setpoint value.
PROP BAND
14 English
Description
The ACTIVATION options change based on the SET FUNCTION setting.
Refer to the tables that follow to configure the analog output.
NONE (default)—Sets the analog output to disabled. [analyzer name]—
Sets the analog output to enabled.
Sets the channel represented at the analog output.
Note: Do not select a channel that contains the symbol "~" (e.g., 4-~SAMPLE4).
Channels that contain the symbol "~" are not measured.
Sets the function of the analog output. Refer to the tables that follow for
more information. LINEAR CONTROL (default)—The analog output is
linearly dependent on the measurement value. PID CONTROL—The
analog output operates as a PID (Proportional, Integral, Derivative)
controller. LOGARITHMIC—The analog output is represented
logarithmically within the measurement range. BILINEAR—The analog
output is represented as two linear segments within the measurement
range.
Sets the value of the analog output when an error occurs if the ERROR
HOLD MODE setting is set to TRANSFER OUTPUTS (0 to 25 mA, default:
4 mA). Refer to
Set the error hold mode
Sets the amount of time for analog output averaging (0 to 999 seconds,
default: 0 seconds). For example, if the value is set to 30 seconds, the
value of the analog outputs is updated every 30 seconds and the value is
the average of the analog output values during the previous 30 seconds.
Description
analog output.
analog output.
Description
AUTO—The analog value (mA) is automatically controlled by the algorithm
when the analyzer uses proportional, integral and derivative inputs. MANUAL
—The analog value (mA) is controlled by the user. To change the value
manually, change the % value in MANUAL OUTPUT.
DIRECT—The analog value increases as the measurement value increases.
REVERSE—The analog value increases as the measurement value
decreases.
Sets a value for the difference between the measured value and the setpoint
value.
on page 22.
Tabla de contenido
loading

Tabla de contenido