4
DISPENSING INTO SAMPLE CONTAINER
Without pressing any key, start dispensing into the
sample container,
Dispensing can be interrupted and started again at will.
Continue dispensing until the level of the fluid in the sample
container has reached the graduated area. There is no need to
reach a preset quantity.
5
SHORT RESET key keying
The meter is informed that the calibration dispensing operation
is finished.
Make sure dispensing is correctly finished before performing
this operation. To calibrate the meter, the value indicated by
the partial totaliser (example 9.800) must be forced to the
real value marked on the graduated sample container. In the
bottom left part of the display an arrow appears (upwards and
downwards), that shows the direction (increase or decrease) of
the value change displayed when the following operations 6 or
7 are performed.
SHORT RESET key keying
The arrow changes direction. The operation can be repeated to
alternate the direction of the arrow.
6
7
SHORT/LONG CAL key keying
The indicated value changes in the direction indicated by the arrow:
- one unit for every short CAL key keying
CAL
CAL
CAL
- continually if the CAL key is kept pressed. The speed increase
rises by keeping the key pressed. If the desired value is
exceeded, repeat the operations from point (6).
8
LONG RESET key keying
The meter is informed that the calibration procedure is
finished.
Before performing this operation, make sure the INDICATED
value is the same as the REAL value.
The meter calculates the new USER K FACTOR ; this calculation
could require a few seconds, depending on the correction to
be made.
ATTENTION: If this operation is performed after action (5),
without changing the indicated value, the USER K FACTOR would
be the same as the FACTORY K FACTOR, thus it is ignored.
12
836 828 R. 10/12
Samoa Industrial, S.A. · Box 103 Alto Pumarín · 33211 Gijón - Asturias Spain · Tel.: +34 985 381 488 · Fax.: + 34 985 384 163
Calibration
Calibration mode
Indicated value
Indicated value
Real value
Real value