AVM52ERT – Analogue Earth Resistance Tester
1. Introduction & Features
Thank you for buying the AVM52ERT ! Please read the manual carefully before bringing this
device into service.
•
The modern design limits the influence of the earth voltage and the earth resistance of
auxiliary earth bars to a minimum.
•
Press the BATT.CHECK button to test the auxiliary earth resistance and the connection
cable. The OK lamp lights up when the device is ready to perform accurate earth resistance
measurements.
•
Low power consumption : max. 12V/100mA.
•
Push buttons allow smooth operation of the device.
•
The earth resistance value can be read directly from the scale.
•
You only need to press the SIMPLIFIED MEAS. button for simplified measurements. No
shorting wire is required since terminals P and C can be shorted internally simply by
pressing this button.
•
Battery replacement is easy.
•
The carrying case in hard plastic is water-resistant and accommodates all accessories.
2. Specifications
Measuring Ranges
Earth Resistance
Earth Voltage
Accuracy
Earth Resistance
Earth Voltage
Measurement System
Measure earth resistance with a CC inverter, 800Hz, ± 2mA
Measure earth voltage through a rectifier, type 5KΩ/V, ± 40~500Hz
Withstand Voltage
Self-Check Facility
Batteries
Dimensions
Weight
Accessories
• test leads (red15m / yellow 10m / green 5m)
• auxiliary earth bars
• carrying case
AVM52ERT
10/100/1000Ω
30V AC (5KΩ/V approx.)
±5% of full scale
±5% of full scale
1500V AC for 60 seconds between electrical circuit and the housing.
Press the BATT.CHECK button to test the earth resistance and the
connection with terminals P and C. The OK lamp comes on.
8 x 1.5V AA-battery
140 (L) x 140 (W) x 90 (D) mm
±800g
1
VELLEMAN