Electromagnetic emission

Conducted RF emission
Test standard
EN 55011
Radiated RF emission
Test standard
EN 55011
Harmonic current emissions
Test standard
IEC 61000-3-2
Voltage changes, voltage fluctuations and flicker
Test standard
EN 61000-3-3
The product must not send too much noise into the
grid . This could lead to a disruption of other products .
Emission test class and group
Class B, group 1
The product must not emit RF signals in the room . This
could lead to a disruption of other products .
Emission test class and group
Class B, group 1
The product must not unintentionally draw power from
the grid .
Emission test class and group
Class A
The product must not affect other products when it is
connected to the grid . For example, these could be lights
that could start flashing and thus bother people in the
room .
Emission test class and group
Tested according to the standard .