• Use of accessories, transducers and cables other than those specified or
provided by the manufacturer could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment
and result in improper function .
• Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than
30 cm (12 inches) to any part of the unit, including its cables, to avoid
any degradation of the performance of this unit .
• Use of this equipment adjacent to or stacked with other equipment
should be voided because it could result in improper operation . If such
use is necessary, this equipment and the other equipment should be
observed to verify that they are operating normally .

Electromagnetic immunity

Electrostatic discharge
en
Test
standard
EN 61000-4-2
Radiated RF electromagnetic field
Test
standard
EN 61000-4-3
20
WARNING
!
The product must not be affected by electrical discharge
between a person and the product in such a way as to
initiate actions or render components defective .
Immunity test level
+/- 8 kV contact
+/- 2, 4, 8, 15 kV air
The product must not be affected by radiating RF signals,
for example from a mobile phone or radio, that could
cause the product to take action or to make components
defective .
Immunity test level
15 V/m 80 MHz to 1000 MHz (only on the front)
15 V/m 1 .01 GHz to 3 GHz
80 % AM 1 kHz