5. Start a tested program with the following properties:
– Sufficient cleaning effect.
– Thermal disinfection: 5 to 10 minutes at 90 to 95 °C,
A
≥ 3000.
0
– Final rinse with distilled or deionized water.
– Sufficient product drying (no visible fluid residues).
6. If contaminants are still visible on the product after the end
of the routine, repeat precleaning and machine cleaning and
disinfection.
7. Disconnect the irrigation tube from the shaft tube.
7.8
Inspection and oiling
1. Check the product for visible damage and wear:
– Damage to the product, e.g. cracks, rough surface, splin-
tering.
– Damage to the product insulation and/or cable/connec-
tor insulation, e.g. cracks and breaks.
– Damage to the coating.
– Bending of the product or parts of the product. Do not at-
tempt to straighten bent instruments!
If highly alkaline cleaners are used, metal surfaces may be dis-
colored. However, this does not affect the integrity of the prod-
uct.
2. Check the mechanical functionality of the product:
– Can the jaws be opened and closed?
– Can the handle be opened and closed?
– ERGO handle (20195-240): Can the upper part of the
handle be rotated freely relative to the lower part? Can
the locking knob be locked and loosened?
If damaged, do not use this product!
3. Oil all moveable parts of the handle and the joint area of the
insert using a physiologically acceptable oil (e.g. paraffin oil).
31