Never lay this product on the patient or in his/her direct vicinity.
Damp inside the connectors may cause flash-overs that could
damage the connector or its insulation. Ensure that the product
is dried sufficiently during processing and prevent fluids from
penetrating the connector during use. The older the material, the
greater the risk of unwanted damp in the connectors.
Erbe Elektromedizin expressly warns against modifying the ca-
ble, e.g. independently replacing defective insulation. Any mod-
ification leads to the termination of liability on behalf of Erbe
Elektromedizin.
4
How to Use
Clean/disinfect and sterilize this product before using it for the
first time and before every subsequent use.
WARNING! Do not kink the cable.
WARNING! Protect this product from any form of mechanical
damage! Do not throw! Do not use force!
WARNING! Risk of injury! Do not activate the electrosurgical unit
while you are connecting the cable.
Connect the instrument to the cable. Then plug the connector on
the cable into a suitable socket on the electrosurgical unit.
Select an appropriate unit setting on the electrosurgical unit. The
unit setting must not exceed the maximum electrical capacity of
both the cable and the instrument as this could damage the ac-
cessories. Information on selecting appropriate unit settings is
provided in the User Manual of the electrosurgical unit.