Civco AccuCARE MICRO-TOUCH STABILIZER Manual De Instrucciones página 6

Tabla de contenido
English
3. Check to see all FAM controls are set near midpoints of their range of travel.
I. Knob YR - Rotates probe to parallel with prostate centerline
II. Knob ZR - Levels prostate horizontally on transverse image
III. Stepper Quick Connect Knob - Secures stepper to FAM
IV. Stabilizer interface plate
V. Knob X - Centers probe along midline axis of the prostate and centers the transverse image on grid
VI. Knob XR - Tilts probe parallel to the posterior surface of the prostate in the sagittal plane
VII. Knob Z - Used to "zero" the stepping unit at the bladder neck to set the base plane for implantation
VIII. Knob Y - Adjusts image height on grid to define posterior row for seed placement
NOTE: Knob Y controls vertical movement and has a useful range of about 6cm (3 up and 3 down). Going down it will bottom out as lower plates come
into contact and it is shipped in this position. To setup for use, turn Knob Y until lower plates are separated by 2-3cm.
ADJUSTING THE STABILIZER
1. Grasp handles of FAM and loosen locking control knob on stabilizer.
®
For Micro-Touch
Stabilizer:
Adjust weight by turning weight adjustment knob on stabilizer to achieve desired feel to equipment.
2. Obtain satisfactory transducer position and starting image, tighten locking control knob on stabilizer. After stabilizer is locked into position, locking
control knob should not be used until removal from the patient.
3. Use FAM to optimize position of probe.
4. Install sterile drape and grid, position grid platform appropriately and secure with grid rail locking knobs. Refer to Sterile Drape and Template Grid
instructions for use.
FUNCTIONAL PREUSE CHECKS
WARNING
Prior to each use perform the following checks to ensure optimal performance.
n
If locking control knob on stabilizer will not hold unit securely in place, please contact CIVCO.
n
1. Attach the table mount at a suitable clasping point on the table rails to ensure it remains securely in position.
2. Ensure stabilizer moves freely when locking control knob is loosened and secures firmly when locking control knob is tightened.
3. Move each of the knobs on the FAM through its full range and then center each knob in position.
4
Micro-Touch® Stabilizer
Tabla de contenido
loading

Este manual también es adecuado para:

Accucare micro-touch stabilizer lp

Tabla de contenido