REPROCESSING
WARNING
Users of this product have an obligation and responsibility to provide the highest degree of infection control to patients, co-workers and
n
themselves. To avoid cross-contamination, follow infection control policies established by your facility.
See your system's user guide for reprocessing transducer between uses.
n
These procedures for reprocessing have been validated for effectiveness and compatibility. Equipment could be damaged or cross-contaminated due
n
to improper reprocessing.
Do not place in a mechanical washer.
n
Do not use ethylene oxide or autoclave to sterilize.
n
CLEANING AND DISINFECTION
1. Using a clean germicidal wipe with at least 55% isopropyl alcohol and 0.5% quaternary ammonium, such as Super Sani-Cloth® Germicidal wipe,
wipe surfaces to remove gross soil.
Ensure all crevices and grooves have come into contact with wipe.
n
2. Using a new, clean germicidal wipe with at least 55% isopropyl alcohol and 0.5% quaternary ammonium, such as Super Sani-Cloth® Germicidal
wipe, wipe surfaces until visibly wet.
3. Allow treated surface to remain wet for no more than two (2) minutes.
Use additional wipe(s) if needed to assure continuous two (2) minute wet contact time.
n
4. Once device has reached two (2) minute wet contact time, allow surface to air dry.
STORAGE AND DISPOSAL
WARNING
Dispose of single-use components as infectious waste. Reprocess reusable components after each use according to instructions for use.
n
NOTE:
Device should be stored to keep all components together and secure.
n
For questions or to order additional CIVCO products, please call +1 319-248-6757 or 1-800-445-6741 or visit www.CIVCO.com.
n
All product to be returned must be in its original packaging. Contact CIVCO for further instructions as needed.
n
8
Micro-Touch® Stabilizer