Replacing the probe base

After several months of use, dust may collect in the probe base, affecting the probe movement. The
probe base should be replaced if the probe no longer moves smoothly.
Unscrew the probe base
collar and put it in a safe
place.

Replacing the batteries

Unscrew the battery compartment locking screw with a screwdriver or a small coin.
Remove the battery compartment cover. Remove the old batteries.
Insert a new set of batteries (four AA batteries). Do not use rechargeable batteries, since they may not
function properly (the inner resistance of some rechargeable batteries is too high). Insert the batteries
in accordance with the diagrams inside the battery compartment, with the +terminals pointing down-
wards on the display side of the tonometer (the rear side), and the -terminals pointing downwards on
the measurement side (the front side).
Replace the battery compartment cover and secure it by screwing it in lightly using the coin or screw-
driver. Take care not to use excessive force when screwing the cover into place.
Remove the probe base
by tilting the tonometer
downwards and pull the
probe base out of the
tonometer.
Insert a new probe base
into the tonometer.
Battery compartment cover
Screw the collar in, to
lock the probe base.