Display after measurements
Before
00
After the sixth measurement, the letter P appears on the display, followed by the IOP (Intraocular
pressure) reading.
If the P is blinking, it means that the standard deviation of the measurements is greater than normal.
P_ (line down) The standard deviation of the different measurements has a slightly greater value than
normally, but the effect on the result is unlikely to be relevant.
P _ (line in the middle) The standard deviation of the different measurements is clearly greater than
normal, but the effect on the result is probably irrelevant. A new measurement is recommended if the
IOP is over 19 mmHg.
_
P
(line up) The standard deviation of the different measurements is great and a new measurement is
recommended.

Other functions

Accessing old measurement values

From the starting position, press the right or left selector button until 'Old' appears on the display. Then
press the measurement button. You can now 'scroll' through the old values by pressing the selector
buttons (right=older, left=more recent, from 0-9).
To exit the old values search, press the measurement button. The display will now show the word 'Old' .
Press either selector button to access other functions (00=measurement, End=turning OFF).

Turning the tonometer OFF

Press either selector button until the display shows 'End'. Press the measurement button for two
seconds - the display will show 'byE' and the tonometer will switch off.
The used probe will be partially ejected. Use the used package to remove it from the tonometer.
Ensure that you dispose of the probe properly.
8
After the second
measurement
2.13
After the sixth measurement.
P 13