QUICK REFERENCE GUIDE: KI7300C / 7700C SERIES OPTICAL LOSS TEST SET
MEASURING ONE WAY LOSS IN AUTOTEST
•
Reference: connect Source port to a Power Meter port. Select [Autotest],
[Abs/Rel]. Press [Set Ref] for 3 seconds. Meter displays about 0.00 dB R.
•
Loss testing: after referencing, connect test system to Source and Power
Meter ports. Loss will be automatically displayed. Use [-/+] to show one
wavelength only.
MEMORY OPERATION
•
Data can be stored in all modes, however, memory recall and cancel
functions are not available in Autotest and only work in Light Source or
Power Meter modes.
•
To clear all memory, press both [MR] and [Exit] simultaneously for 3
seconds. 'clr' will display.
•
To store in next default location, press [M+].
•
To change next store location, press [◄►] [-/+] [Set] or [Menu].
•
To recall, press [MR] and [-/+] to scroll memory.
•
Exit Autotest and enter Source or Power Meter mode to read or cancel
memory.
•
Autotest memory recall: use [-/+] to scroll wavelengths.
•
To exit memory display, press [Menu].